Table des Matières

Publicité

®
R&S
RT-ZD40
To assign "Set offset to mean" to the micro button
► "Vertical" menu > "Probe Setup" > "Ch" tab > "Micro button action" = "Set off-
set to mean"
2.3.2

Common Mode Offset

Common mode offset compensation is available for R&S RT‑ZD probes with
serial number ≥ 200 000. It can compensate a common DC voltage applied to
both input sockets (referenced to the ground socket). This is particularly helpful
for measurements on differential signals with high common mode levels, for
example, current measurements using a shunt resistor.
Fig. 2-3: Common mode (CM) offset compensation for a differential measurement
If the input signals fit into the operating voltage window of the R&S RT-ZD40, it is
not necessary to set a common mode offset compensation.
User Manual 1410.5363.02 ─ 03
Putting into Operation
Offset Compensation
20

Publicité

Table des Matières
loading

Produits Connexes pour R&S RT-ZD40

Table des Matières