To change these values, proceed as follow:
• SpO 2 alarm max. value
Press Select once, the red led (SpO 2 Min) lights up and a value blinks on the left display. This value
represents the maximum value for oxygen saturation (SpO 2 ).
To change this value, press key
• SpO 2 alarm min. value
Press Select two times, the red led (SpO 2 Min) lights up and a value blinks on the right display. This
value represents the minimum value for oxygen saturation (SpO 2 ).
To change this value, press key
• Heart rate alarm max. value
Press Select three times, the red led (Pulse Max) lights up and a value blinks on the left display. This
value represents the maximum value for heart rate (Pulse/min). To change this value, press key
increase it and key
• Heart rate alarm min. value
Press Select four times, the red led (Pulse Max) lights up and a value blinks on the right display. This
value represents the minimum value for heart rate (Pulse/min). To change this value, press key
increase it and key
When it stops blinking, the equipment returns to the original conditions.
The new alarm settings will be automatically saved.
When the batteries run down or when the equipment is disconnected from power supply, the original
alarm settings are restored.
During measurement, the pulse oximeter display shows an intermittent scale of 10 leds indicating the
search for beating. As seconds go by, the scale grows longer and becomes stable. When the signal is
good, the search bar should show at least 4 leds.
If the loud speaker is on, the equipment gives out intermittent audio signals. Press
stop the sound and the red led
starts up again.
The alarm keeps on sounding for a minute and then it stops.
Switching off
After the measurement has been taken, turn the equipment off by means of ON/OFF pushbutton. A
beep is given out and then the equipment stops.
General maintenance instructions
The product has been conceived to last forever; it is resistant to most part of chemicals except for
methyl ethyl ketones (M.E.K.), diluents and nail polish removers.
To clean the instrument, make use of a cloth dampened with disinfectant.
to increase it and key
to increase it and key
to reduce it.
to reduce it.
, on the display lights up. Press the pushbutton again and the sound
MAINTENANCE
to reduce it.
to reduce it.
11
to
to
pushbutton to