4.2.7
Continuity test
Range
Resolution
Uncertainty
Audible signal
triggered
600
0.1
50
± 10
4.2.8
Diode test
Range
Resolution
Uncertainty (±)
2.8 V
0.001 V
1 % + 2 cts
4.2.9
Capacity
Range
Resolution
40 nF
0.01 nF
400 nF
0.1 nF
4 µF
0.001 µF
40 µF
0.01 µF
400 µF
0.1 µF
1000 µF
1 µF
4.2.10
Temperature (type K thermocouple)
The uncertainty does not include the K thermocouple.
Range
Resolution
- 20 to 760 ° C
1° C
- 4 to 1400 ° F
1° F
4.2.11
Max/Min
Capture time: 400 ms
Add an uncertainty of ± (0.5% + 2 pts) to the uncertainty of the
function and of the range used.
English
Measurement
current
< 0.35 mA
Open-circuit
Measurement current
voltage
< 2.8 V
< 0.9 mA
Uncertainty (±)
3.5 % + 4 cts
5 % + 5 cts
Uncertainty (±)
2 % + 5° C
2 % + 9° F