3.0 Adjusting Temperature Settings

The following section describes how to adjust the MVE Variō system temperature and alarm settings. At any
time, the user may exit the current menu by pressing the ESC button repeatedly until the display returns to
the main screen. After 30 seconds of inactivity, the controller will automatically exit the current menu and
return to the main screen.

3.1 Adjusting Storage Chamber Temperature

1. From the home screen, press SETUP to enter setup menus. If Password Entry
Mode is enable, the controller will prompt for a password. Use the
scroll to the appropriate number and press ENTER to advance the cursor to the
next position.
Note: The default user password is 3 4 5 6.
2. After entering password, the first menu item is Temperature Menus. Press
ENTER.
3. Press SETUP to scroll to the Cooling Menu display and press ENTER.
4. The first menu item will display the current temperature of the cooling temp
sensor. Press SETUP to continue.
5. To adjust the maximum (warmest) allowable chamber temperature, use the
▲/▼
keys to select the desired value and press ENTER to save. Press SETUP to
advance to the next screen.
Note: The Variō Pro controller will not allow the user to select a maximum chamber temp greater than
or equal to the high temperature alarm setpoint. As a result, it may be necessary to adjust the high
temperature alarm prior to changing the maximum chamber temp. See the following section
for further instruction.
6. To adjust the chamber temperature deadband, use the
desired value and press ENTER to save. Note that the chamber deadband also
defines the minimum (coldest) allowable chamber temperature.
Note: The Variō Pro controller will not allow the user to select a maximum chamber temp
chamber deadband combination that results in a minimum chamber temp less than or equal to the
low temperature setpoint. As a result, it may be necessary to adjust the low temp alarm prior to
changing the chamber deadband.
7. Press SETUP repeatedly to scroll to the Cooling Time Alarm display. For the
initial cooling cycle, it will be necessary to disable the alarm. Use the
to select DISABLED and press ENTER to save. Press ESC repeatedly to return to
the home screen.
Note: It is recommended that the Cooling Time Alarm be re-enabled after the initial cooling cycle is
complete. Follow the steps above to return to the Cooling Menu and set the Cooling Time Alarm to
ENABLED.
PN 14910831 Rev A
MVE Variō™ Pro Quick Start Guide
Higher User Level
▲/▼
keys to
Required use
ENTER Password
0000
Press ENTER for
Temperature menus
or press SETUP for
next menu
Press ENTER for
Cooling menu
or press SETUP for
next menu
Current Cooling Temp
-60.0°C
Max Chamber Temp
-80.0°C
Use
Press ENTER to save
▲/▼
keys to select the
Chamber Deadband
10.0°C
Use
▲/▼
Press ENTER to save
Cooling Time Alarm
▲/▼
keys
ENABLED
Use
Press ENTER to save
▲/▼
to
▲/▼
to adjust
▲/▼
to adjust
▲/▼
to adjust
Page 5