Calibration Of Temperature Probes - CHART BioMedical MVE Vario Pro Guide De Référence Rapide

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

MVE Biological Systems
MVE Variō™ Pro Quick Start Guide
TM

5.0 Calibration of Temperature Probes

Each of the temperature sensor probes used with the Variō Pro have been calibrated at the factory using the
Low Temperature Range method. This calibration method provides a level of accuracy of +/-1.8°F or +/-1°C
when operated at an altitude range of 1000ft – 1500ft (305m - 457m). Further calibration should not be
required unless desired by the user. Refer to the Variō Pro Technical Manual (PN 14930250) for information
on calibration methods and procedures.
Page 9