Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

RIKTA
- VET
®

MAINTENANCE

In order to ensure reliable operation of the Device it is necessary to
perform the proper maintenance as specified in Table 2 hereof.
During the maintenance, make sure to take safety measures specified
in section "SAFETY MEASURES", page 24.
If the Device is found non-conforming to the technical requirements
specified in Table 2, it may not be operated any more and is subject to
repair.
Maintenance
Frequency
Before each
session
Work Scope
Visual inspection not
using any special tools
Check the following:
1. State of battery charge
by Battery indicator [3]
2. Availability of radiation
at the emitting window
output [9]
3. Disinfection
(if required)
Technical Requirements
No mechanical damages
on the housing of Device
and battery charger
Battery indicator [3]
should be off
Once the emitting win-
dow [9] of the Device is
targeted at the exposure
area and START/STOP but-
ton is pressed, visible red
light radiation should be
observed
According to
recommendations given
in section "SAFETY
MEASURES", page 24
35
Table 2
EN

Publicité

Table des Matières
loading

Table des Matières