Timeout
This feature allows the user to preset an automatic switch off when not in use of between 1-
10 minutes.
Sound
The speaker can be switched on or off. If switched off both the menu selection confirmation
and the acoustic positioning guidance sound will be disabled.
Units
User selectable between microns (µm) or mils.
2.6.7
Limits
Set up
The user can set high and low predicted cured thickness limit boundaries for their quality
control process. If a measurement is taken which falls outside these limit boundaries then the
gauge displays a warning error message.
In addition to setting limits the Elcometer 550 Non-Contact Uncured Powder Gauge features
powder shrinkage compensation. This adjustment is known as the percentage distortion
setting and allows users to ensure that the predicted coating thickness value is equivalent to
the actual coating thickness post cure.
To calculate the percentage distortion value:
Measure a powder coated test panel using Elcometer 550 Non-Contact Uncured Powder
Gauge and record the gauge's predicted coating thickness value. Cure the test panel in
the powder oven for the appropriate time and temperature for the powder coating
R