Accessing parameters via the handheld programmer
: SITRANS LR250 automatically returns to Measurement mode after a period of
Note
inactivity in PROGRAM mode (between 15 seconds and 10 minutes, depending on the menu
level).
Parameter menus
Parameters are identified by name and organized into function groups, then arranged in a 5-
level menu structure. For the complete list of parameters with instructions, see the full manual.
Note:
In Navigation mode
ARROW keys move to the
next menu item in the
direction of the arrow.
1.
Enter PROGRAM mode
• Point the programmer at the display (from a
maximum distance of 600 mm [2 ft.]).
• Right ARROW
and opens menu level 1.
• Mode
opens the menu level last displayed
in PROGRAM mode within the last 10 minutes, or
menu level 1 if power has been cycled since then.
2.
Navigating: key functions in Navigation mode
Key
Name
Up or Down
ARROW
Right ARROW
Left ARROW
Mode
Home
Page EN-14
SITRANS LR250 (HART) – QUICK START MANUAL
activates PROGRAM mode
Menu level
menu or parameter
menu
parameter
menu or parameter Open parent menu.
menu or parameter Change to MEASUREMENT mode.
menu or parameter Open top level menu: menu 1.
1. QUICK START
2. SETUP
2.2. INPUT
2.2.5. TVT SETUP
2.2.5.6. AUTO ECHO SUPP
handheld programmer
(ordered separately)
Max. 600 mm
(2 ft)
Function
Scroll to previous or next menu or
parameter.
Go to first parameter in the
selected menu, or open next menu.
Open Edit mode.
display
7ML19985QX81