A 9-volt battery powers the Oxford Electronic Digital Scale. A new battery should provide approximately 3000
readings before needing replacement. When it needs to be replaced, a "B" will appear on the display.
To Replace Battery:
1. Push black pad on bottom of unit to left.
2. Battery holder will slide out.
3. Remove old battery and replace with a new one.
4. Push holder back into scale.
OPERATION FEATURES:
• All functions are accessible to the user without opening the case.
• Starting or zeroing the scale is done through a single touchpad located on the front of the case.
• Stable readings are achieved through digital averaging. The HOLD message is activated when
weight fluctuations decrease to a predetermined range. The scale becomes active again when that
range is exceeded. The held weight is stored in a non-volatile memory for later recall if required.
• Display turns off after 60 seconds of inactivity to conserve battery life.
• Recall of the last held weight is available via a touchpad on the front of the case. This allows a
reading of the last patient's weight, after the patient has been removed from the scale.
• Battery change is made easy with external access to the battery holder.
MAINTENANCE, INSPECTION AND TEST:
Joerns Healthcare recommends that a thorough inspection and test of all Oxford lifts and lifting acces-
sories, slings etc. is carried out every six months. Joerns Healthcare recommends maintenance, inspec-
tion and certified testing, is only carried out by competent and authorised service dealers.
Note: These recommendations are in compliance with the requirements of 1998 No2307 Health and
Safety: The Lifting Operations and Lifting Equipment Regulations 1998. This is a UK regulation.
Outside the UK please check your local requirements.
Important Note (Europe Requirement Only)
The Oxford Weigh scale is a class 3 compliant weighing device and conforms to the NAWI Directive
90/384/EEC and has been certified by the National Weights and Measures Laboratory.
NOTE: Scale is calibrated prior to shipment.
This certifies the scale for twelve months. Joerns Healthcare's advice is that you contact your local
trading standards / government authority or Joerns Healthcare for your nearest authorised body who
can advise upon recalibration.