1.0 Indications for Use

This device is intended to aid in the prevention of
inadvertent hypothermia during administration of
blood, blood products, and other fluids.

1.1 Safety Precautions

DANGER
Risk of electric shock.
No user serviceable parts inside.
WARNING
• All air must be removed from the fluid lines
prior to connection to the patient.
• Monitor the fluid lines to insure they are air
free. Never administer fluids if there are air
bubbles in the line between the bubble trap
and the patient connector.
• Monitor the fluid level in the bubble trap
frequently. Bubbles released during heating
are captured in the trap. To refill the trap,
insert a sterile syringe into the valve and draw
air out until the trap is (2/3) full. Do not allow
the bubble trap to go below one-quarter
(1/4) full.
• The bubble trap must be kept mounted and
in the vertical position at all times.
Failure to follow the above warnings
could result in the introduction of air to
the patient. Introduction of air to the
patient could result in death or serious
injury.
CAUTION
• US Federal law restricts this device to sale by
or on the order of a physician.
• Do not immerse in cleaning and/or
sterilization solution. Do not submerge or
soak unit; it is fluid resistant, not fluid proof.
• For grounding reliability of the Blood/Fluid
warmer, plug only into a properly grounded
outlet.
• The RUN/STANDBY switch does not provide
isolation from the mains. Isolation from the
mains (IEC 601-1) can only be achieved by
disconnecting the cord from the mains.