Textron Greenlee DM-860A Manuel D'instructions page 18

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Accuracy
Refer to the "Specifications" section for operating conditions and temperature coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a fixed amount) at 23 °C ± 5 °C
(73.4 °F ± 9 °F), 0% to 75% relative humidity. Specifications are for 50,000 counts mode.
True RMS Readings: Voltage and current accuracies are specified from 5% to 100% of the range
unless otherwise specified. Frequency must be within the specified bandwidth for non-sinusoidal
waveforms. Crest factors are as follows:
• Crest factor < 2.1:1 at full scale
• Crest factor < 4.2:1 at half scale
AC, DC
AC
, and AC + DC
Accuracy at DC
Range
and 50 Hz to 60 Hz
500.00 µA
± (0.5% + 0.5 µA)
5000.0 µA
± (0.5% + 5.0 µA)
50.000 mA ± (0.5% + 0.05 mA) ± (0.7% + 0.05 mA) ± (2.0% + 0.05 mA) ± (5.0% + 0.05 mA)
500.00 mA ± (0.5% + 0.5 mA)
5.0000 A
± (0.5% + 0.005 A) ± (0.7% + 0.005 A)
10.000 A*
± (0.5% + 0.05 A)
*
10 A continuous; 20 A maximum (Duty Cycle: 30 seconds on, 5 minutes off).
Burden Voltage
Range
(all frequency ranges)
500.00 µA
5000.0 µA
50.000 mA
500.00 mA
5.0000 A
10.000 A
DC Current
Range
Accuracy
500.00 µA
± (0.15% + 0.2 µA)
5000.0 µA
± (0.1% + 2.0 µA)
50.000 mA
± (0.15% + 0.02 mA)
500.00 mA
± (0.15% + 0.3 mA)
5.0000 A
± (0.5% + 0.002 A)
10.000 A*
± (0.5% + 0.02 A)
*
10 A continuous; 20 A maximum (Duty Cycle: 30 seconds on, 5 minutes off).
18
AC
Current
Accuracy at
40 Hz to 1 kHz
± (0.7% + 0.5 µA)
± (0.7% + 5.0 µA)
± (0.7% + 0.5 mA)
± (0.7% + 0.05 A)
0.15 mV/µA
3.3 mV/mA
45 mV/A
Burden Voltage
Accuracy at
1 kHz to 20 kHz
± (2.0% + 0.5 µA)
± (2.0% + 5.0 µA)
± (2.0% + 0.5 mA)
Unspecified
Unspecified
0.15 mV/µA
3.3 mV/mA
45 mV/A
Accuracy at
20 kHz to 100 kHz
± (5.0% + 0.5 µA)
± (5.0% + 5.0 µA)
± (5.0% + 0.5 mA)
Unspecified
Unspecified

Publicité

Table des Matières
loading

Table des Matières