Xylem SI Analytics ViscoPump II Mode D'emploi page 30

Appareil de mesure de la viscosite
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Initial operation
2.3.2 Deployable hose combinations
Hose combinations
Type no.
VZ 5505
VZ 5505 + VZ 8526
VZ 8523
VZ 8524
VZ 8530
VZ 5606
Note
The hose combinations are to be selected on the basis of the required application,
e.g. VZ 8521, VZ 8523, and VZ 8524.
Note
When using Micro Ubbelohde viscometers with TV sensors, the ignition temperature of the
media to be measured has to be taken into account:
− It has to be higher than 250 °C.
30
Description
Silicone hose kit, oppressive,
for Ubbelohde viscometers
(3 legs), and Cannon-Fenske
and Ostwald viscometers
Silicone hose kit, suctioning,
for Ubbelohde viscometers
(3 legs)
PTFE hose set, suction,
for Ubbelohde viscometer
(3 legs)
PTFE hose kit, suctioning,
with soda lime filter VZ 7215
for Ubbelohde viscometers
(3 legs)
PTFE hose kit, suctioning,
with soda lime filter VZ 7215
for Ubbelohde viscometers
(4 legs)
For TC viscometers (3 and 4 legs)
with screw connections: Silicone
hose fittings with connecting
cable. For oppressive operation
Application
Standard, but please observe:
the sample can leak from the capillary tube
during a malfunction
Standard, safer than oppressive operation,
as the sample cannot leak from the capillary
tube. Unsuitable for volatile samples.
For aggressive samples that attack silicone,
such as sulphuric acid. Hose lengths are
designed to fit the AVS®370 on the support
console VZ 8571. All hose sets for suctioning
operation can be combined with the suction
set and the sample fill set.
For aggressive samples whose vapors are
absorbed by soda lime filters to protect the
ViscoPump. Depending on the sample, an
active carbon filter VZ 7216 can be used
instead of the soda lime filter VZ 7215.
Hose lengths are designed to fit the AVS®370
on the support console VZ 8571.
For Ubbelohde viscometers with additional
4th pipe for filling and cleaning Hose
lengths are designed to fit the AVS®370
on the support console VZ 8571.
For TC viscometers (3 and 4 legs).
Typical applications are measurements of oils.
ViscoPump II

Publicité

Table des Matières
loading

Table des Matières