Health O Meter 600KL Instructions D'utilisation page 12

Masquer les pouces Voir aussi pour 600KL:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

OPERATING INSTRUCTIONS (CONTINUED)
Using the Scale for Height Measurement Only
1. Press the ON/OFF button to turn the scale on.
2. Wait until '0.0' and 'Zero' appear on the left side of the display.
3. Press the WEIGHT/HEIGHT/BMI button once. The scale will display the position of the height
rod.
4. Ask the patient to step on the scale platform. Adjust the height rod so that the headpiece gently
rests on the patient's head. Read the height.
5. Press the RESET button to revert back to weight display.
To Obtain BMI
1. Press the ON/OFF button to power on the scale.
2. Ask the patient to step on the scale platform and obtain the patient's weight. Press the
WEIGHT/HEIGHT/BMI button once. The scale will now display the position of the height rod.
While the patient remains on the platform, adjust the height rod so that the headpiece gently rests
on the patient's head. Press the WEIGHT/HEIGHT/BMI button again. The display will show the
patient's BMI.
3. If desired, press the SEND button to send the weight, height and BMI information to the scale
USB port.
4. Press the RESET button to revert back to weight display.
600KL KG Only Procedure
Caution: Please make sure you carefully and precisely follow the directions below as you will
be in the PROGRAM MODE of your scale. Any modification outside of what is listed below
may greatly affect the functionality of your scale.
1. Press the ON/OFF button to power on the scale. As soon as you turn on the scale, press the
ZERO button once, the UNIT button twice, and then the ZERO button once again. The scale will
display "CAL".
2. Press the
once and "PROG" will show on the display.
3. Press the SEND button and OP1-1 will be displayed.
4. Press the
once and the display will read "OP2-0"for the KG only function, then press the UNIT
button to enable KG only mode. Your display should read "OP2-1".
5. Press the
(7 times) until display reads "DONE"
6. Press the SEND button, the scale will read "START", then read "0.0 Kg"
You have successfully locked your scale into KG Only.
600KL KG Only (Deactivation) Procedure
1. Press the ON/OFF button to power on the scale. As soon as you turn on the scale, press the
ZERO button once, the UNIT button twice, and then the ZERO button once again. The scale will
display "CAL".
2. Press the
once and "PROG" will show on the display.
3. Press the SEND button and OP1-1 will be displayed.
4. Press the
once and the display will read "OP2-1"for the KG only function, then press the UNIT
button to disable KG only mode. Your display should read "OP2-0".
5. Press the
(7 times) until display reads "DONE"
6. Press the SEND button, the scale will read "START", then read "0.0"
You have successfully deactivated the KG Only feature on your scale.
12

Publicité

Table des Matières
loading

Table des Matières