Cleaning And Disinfection - VARILITE Icon Back Low Instructions D'installation

Système de support lombaire
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 11
Icon
Back System / JUNIOR Back Support Installation Manual
EN
3. Remove the back cushion from the attached flap on the inside of the cover.
Note: Installation is the reverse of removal.
ADDITIONAL BRACKET CONFIGURATIONS
The hardware is designed to support alternate configurations, shown below
Icon Back
BASIC
EXTRA DEPTH
JUNIOR Back Support
SPACE SAVER
GROWTH
TROUBLESHOOTING
For additional assistance, please contact VARILITE customer service.
The VARILITE Back System should operate smoothly. The back support should remove and insert without binding. If
the back support binds or seems "sticky" during removal and insertion, quick and easy adjusments will correct the
situation.
Only loosen the Cane Bracket Screw three-quarter (3/4) turn when making depth or angle adjustments.
Loosening the screw more than a three-quarter (3/4) turn will cause it to disconnect from the nut inside the Holster.
1. Check the back height. Ensure that the Cane Brackets are installed at the same height and are parallel.
2. Allow the Slide Brackets to self-center. Loosen the left and right Slide Bracket screws until the Slide Brackets fit loosely
and self-center in the Holsters. Tighten the Slide Bracket Screws.
3. Check the depth adjustment. Ensure that the depth adjustment is the same on the left and right Holsters. Loosen
the top and bottom Cane Bracket Adjustment Screws three-quarter (3/4) turn. While holding the back system at the
appropriate depth, tighten the top and bottom adjustment screws.
4. Check the angle adjustment. Ensure that the angle adjustment is the same on the left and the right Holsters. Loosen up
the top Cane Bracket Adjustment Screws a three-quarter (3/4) turn. While holding the back support at the appropriate
angle, tighten the top adjustment screw.

CLEANING AND DISINFECTION

Warning: Do not use quaternary ammonium-, chlorine- or hydrogen peroxide-based products to disinfect your back
support. These products can damage your back support.
Disinfectants are NOT effective on porous surfaces such as foam or wood. If necessary, they may be cleaned by hand
wiping with a damp cloth. They cannot be effectively disinfected and must NOT be used by different individuals.
VARILITE recommends a minimum 70% solution of isopropyl alcohol for use as a disinfectant.
When cleaning and disinfecting the shell, avoid getting liquid underneath the edge trim.
Cover Cleaning and Disinfection Procedure:
See laundering symbols in this document and on care and use tag.
Cushion Cleaning and Disinfection Procedure:
1. Hand wash cushion under hot tap water (40C) until all visible soil is removed from the cushion. Use mild liquid dish
soap if needed. Repeat if necessary.
2. Rinse the cushion thoroughly, so that the soap is completely removed from the cushion.
3. Let cushion dry for at least 30 minutes.
4. Spray entire cushion with 70% Isopropyl alcohol (IPA) ensuring that the entire pad is saturated.
5. Let cushion dwell with 70% IPA for 10 minutes.
6. Repeat steps four and five once more.
Ensure cushion is fully dry before resuming normal use.
Shell Cleaning and Disinfection Procedure:
1. Hand wash shell with hot tap water (40C) until all visible soil is removed. Use mild liquid dish soap if needed.
Repeat if necessary.
2. To disinfect, spray with 70% Isopropyl Alcohol, ensuring all surfaces are saturated. Maintain a 10-minute wet time.
Ensure shell is fully dry before resuming normal use.
:
GROWTH
EXTRA GROWTH
ADJUST
ANGLE
ADJUST
WIDTH
Cover Cleaning
and Disinfecting
Instructions
Cushion and
Shell Cleaning
Instructions
Cushion and
Shell Disinfecting
Instructions
©2021 Cascade Designs, Inc. #34-287

Publicité

Table des Matières
loading

Produits Connexes pour VARILITE Icon Back Low

Table des Matières