Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 87

12. TECHNICAL DATA

CEFAR BASIC is a nerve stimulator (TENS) with two non-independent channels and three preset pro-
grams.
Treatment with electrical stimulation requires the stimulation current to penetrate the resistance of
the skin and the electrode, about 1500 ohms. CEFAR BASIC can penetrate this resistance and maintain
a current of up to 60 mA. With a change in load from 100 to 1500 ohms, the stimulation current chan-
ges less than 10% from the set value.
The stimulator operates on either two non-rechargeable 1.5 V AA batteries or two rechargeable 1.2 V
AA batteries, recharged in a separate charger.
CEFAR BASIC
Number of channels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 (non-independent)
Constant current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Up to a resistance of 1500 ohm
Stimulation current/channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0–60 mA (charge maximum: 10.8μC)
Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Asymmetrical biphasic pulse, 100% compensated
Number of preset programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Stimulation forms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Conventional
Max pulse duration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180 μs
Max frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Hz
Timer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 60 min
Environment for
storage, use and shipping. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Temperature 10ºC–40ºC
Power source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 x 1.5 V AA non-rechargeable or
Current consumption for
one channel, 80 Hz, 30 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150 mA
I r.m.s. max/channel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 mA
Size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 x 50 x 30 mm
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ca. 180 g
(increased load can reduce the maxiumum current)
Burst
Modulated pulse duration
Air humidity 30%–75%
Air pressure 700 hPa–1060 hPa
2 x 1.2 V AA rechargeable batteries