Press the memory button
screen. The average value of morning measure-
ments of the last 7 days is displayed (morning:
5 am – 9 am).
Press the memory button
screen. The average value of evening measure-
ments of the last 7 days is displayed (evening:
6 pm 8 pm).
If you press the memory button again
again, the latest measurement is displayed on
the screen. The irregular heartbeat symbol (if
there is one)
and the symbol for ranking
(17)
the measuring results
(18)
If you press the memory button
see your individual measurements.
To switch the device off, press the STOP/START
button
.
(2)
You can leave the menu at any time by pressing
the STOP/START button
(2)
8
/ 74
ABPM-100 EN Instruction manual
Deleting the values measured
. A7 flashes on the
(3)
ERROR MESSAGE/RECTIFYING ERRORS
Problem
The LCD screen
displays an
abnormal result.
. P7 flashes on the
(3)
(3)
·
flash.
·
·
·
again, you can
·
(3)
·
·
.
To delete the records of a specific user memory,
first select a user memory.
Hold the memory button
down for 3 seconds
(3)
to delete the results.
Cause
The cuff is in the
wrong position or has
not been tightened
sufficiently.
The body's posture
is not correct
during the test
Arm or body move-
ment, nervousness,
excitement or the act
of talking during the
measuring process
Heart rate is irregular
(arrhythmia)
In the event of error, an error message is displayed
on the screen Er_.
Error messages may be displayed when:
the systolic or diastolic pressure has not been
measured (ER1 or ER2 is displayed on the screen);
systolic or diastolic pressure is outside the
measuring range ("Hi" or "Lo" is displayed on
the screen);
the cuff is too tight or too loose (ER3 or ER4 is
displayed on the screen);
blood pressure is higher than 300 mmHg (ER5 is
displayed on the screen);
inflation takes more than 180 seconds (ER6 is
displayed on the screen);
there is an error on the system or the device
(ER7 or ER8 or ERA is displayed on the screen);
the batteries are nearly flat
(18)
Solution
Apply the cuff cor-
rectly and try again.
Re-read sections
"Adjusting to the
correct position"
in the instructions
and try again.
Take another
measurement and
follow the positioning
instructions and
instructions for use.
The blood pressure
monitor is not
suitable for people
suffering from
serious arrhythmia
.