Laboratory pump Laboport® N96
Operating
parameters and
conditions
Requirements on
the transferred
medium
Accessories Laboratory accessories or additional components that are
Translation of Original Operating Instruction, english, KNF
323241-323242 06/20

3 Use

3.1 Proper use

The pumps are intended exclusively for transferring gases
and vapors.
Responsibility of the owner
Only install and operate the pumps in accordance with the op-
erating parameters and conditions described in Chapter 5
Technical data.
Only pumps that are fully assembled and in the condition as
delivered may be operated.
Make sure that the installation location is dry and that the
pump is protected against rain, splash, gushing, and drip wa-
ter as well as from other contaminants.
Check the tightness of the connections between the pipes of
the application and the pump (or the connection of the pump)
at regular intervals. Leaky connections carry the risk of re-
leasing dangerous gases and vapors from the pump system.
Before transferring a medium, check whether the medium can
be transferred danger-free in the specific application.
Before using a medium, check the compatibility of the media-
contacting components (see 5 Technical data) with the
medium.
Risk of dangerous gas mixtures during pump operation if di-
aphragm breaks: Depending on the medium being trans-
ferred, breakage of the diaphragm can result in a dangerous
mixture if the medium mixes with the air in the compressor
housing or the surroundings.
Only transfer gases that remain stable under the pressures
and temperatures that arise in the pump.
connected to a pump must be designed for the pneumatic
data of the pump (see 5 Technical data).