Ets-D Or Contact Thermometer; Setting The Temperature Control Mode; Maintenance; Accessories - IKA RCT basic Mode D'emploi

Masquer les pouces Voir aussi pour RCT basic:
Table des Matières

Publicité

Setting the temperature control mode

When using an external PT 1000 temperature sensor, the user can
choose between two types of control:
PI mode
Good control results, minimized overshooting, slow rise in temperature
2P-mode (two-point controller)
Maximum heating rate, increased overshooting
Factory setting: PI mode
Changing the temperature control mode
1. Set target temperature to 2 °C
2. Switch off device
3. Hold down temp knob
4. Switch on device
This activates or deactivates the two-point controller (2P) for the
external PT 1000 temperature sensor (2P appears on the display).
When switching on the device, 2P is also displayed next to the
operating mode A/B/D when the two-point controller is activated.
When the heating function is active in 2P mode, (with external PT 1000
temperature sensor) the actual temperature flashes - that notifies the
user that the process temperature has been greatly overshot!

Maintenance

The device is maintenance-free.
Cleaning
For cleaning disconnect the main plug.
Only use cleansing agents which have been recommended by IKA
Use to remove:
Dyes
isopropyl alcohol
Construction materials isopropyl alcohol/water containing surfactant
22
Cosmetics
isopropyl alcohol/water containing surfactant
Foodstuffs
water containing surfactant
Fuels
water containing surfactant
- Do not allow moisture to get into the appliance when cleaning.
- Wear protective gloves when cleaning the devices.
- Please consult with IKA
®
mination methods, other than those recommended here.
Ordering spare parts
When ordering spare parts, please give:
- Machine type
- Manufacturing number, see type plate
- Item number and designation of the spare part,
see www.ika.com.
Repair
Please send instrument in for repair only after it has been cleaned
and is free from any materials which may constitute a health hazard.
For this you should request the "Decontamination Clearance Cer-
tificate" from IKA
, or use the download printout of it from the
®
IKA
website www.ika.com.
®
Return the instrument in its original packaging. Storage packaging is
not sufficient. Also, please use suitable shipping package materials.

Accessories

• Stirring bars: ø 6 mm, length up to 15 mm
ø 7 mm, length up to 60 mm
ø 10 mm, length up to 80 mm
• RS 1
Set of magnetic stirring bars
• RSE
Stirring bar remover
• H 15
Bath attachment
• H 29
Oil bath attachment
®
.
• H 16 V Support rod
• H 38
Holding rod
• ETS-D5 Contact thermometer
• ETS-D6 Contact thermometer
before using any cleaning or deconta-
• H 28
Bath attachment
• H 30
Oil bath attachment
• H 16.3 Extension
• H 44
Cross sleeve