35
Recommended separation distances between portable and mobile RF communications equipment and the
EQUIPMENT or SYSTEM – for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and
the Spirometer
The Spirometer is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the Spirometer can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the Spirometer
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Separation distance according to frequency of transmitter (m)
80 MHz to 800 MHz
3.5
d=
E
1
0.12
0.37
1.17
3.69
11.67
ENGLISH
√P
800 MHz to 2,5 GHz
7
√P
d=
E
1
0.23
0.74
2.33
7.38
23.33