Specifications
Display:
9999 counts: ACV, DCV, Hz, and nS
6000 counts: mV, µA, mA, A, ohm, and capacitance
Polarity: Automatic
Sampling Rate:
Numeric Display: 5 per second
41-Segment Bar Graph Display: 60 per second
Temperature Coefficient: Nominal 0.15 x (specified accuracy) per °C
below 18 °C or above 28 °C
Automatic Power Off: After 30 minutes of inactivity.
To disable this feature, press SELECT while turning the meter on.
Noise Rejection*:
Normal Mode Rejection Ratio > 60 dB at 50 Hz and 60 Hz when measuring DCV
Common Mode Rejection Ratio > 60 dB from 0 Hz to 60 Hz when measuring ACV
Common Mode Rejection Ratio > 120 dB at 0 Hz, 50 Hz and 60 Hz when measuring DCV
Operating Conditions:
0 °C to 45 °C (32 °F to 113 °F), 0% to 80% relative humidity (non-condensing)
Altitude: 2000 m (6500') maximum
Indoor use only
Pollution Degree: 2
Storage Conditions: –20 °C to 60 °C (–4 °F to 140 °F)
0% to 80% relative humidity (non-condensing)
Remove battery.
Battery: 9-Volt (NEDA 1604, JIS 006P or IEC 6F22)
Overload Protections:
DM-810A, DM-820A, and DM-830A:
Volts: 1050 V RMS AC/DC, 1450 V peak AC/DC
Millivolts: 600 VDC and VAC RMS
A: 11 A / 1000 V fuse, interrupting rating 20 kA, F fuse, 13/32" x 1-1/2"
µA and mA: 0.44 A/1000 V fuse, interrupting rating 10 kA, F fuse, 13/32" x 1-1/2"
Other Functions: 600 VDC and VAC RMS
DML-430A:
Volts: 1050 V RMS AC/DC, 1450 V peak AC/DC
Millivolts: 1050 V RMS AC/DC, 1450 V peak AC/DC
A: 11 A/1000 V fuse, interrupting rating 20 kA, F fuse, 13/32" x 1-1/2"
µA and mA: 0.44 A/1000 V fuse, interrupting rating 10 kA, F fuse, 13/32" x 1-3/8"
Other Functions: 1050 V RMS AC/DC, 1450 V peak AC/DC