25. Do not attempt to stop the rotor while unit is running. Doing so may cause the unit to fail and
will void the warranty.
26. Only use original manufacturer's rotors, AC adapters, tube adapters and any other spare parts.
27. Repairs are to be performed by trained and authorized service personnel only. Contact Thermo
Fisher Scientific service.
28. Opening the instrument housing or improper use of the instrument voids the warranty. If there
is a failure during the warranty period, contact Thermo Fisher Scientific service for warranty
service.
29. Do not attempt to use units that have not been correctly installed or repaired.
30. Do not attempt to disassemble or modify this product.
31. Store at room temperature in a dry area. Do not expose to sunlight, moisture, or extreme tem-
peratures for prolonged periods of time.
32. Before using any cleaning or decontamination methods except those recommended in this
manual, please contact Thermo Fisher Scientific to ensure that the proposed method will not
damage the equipment.
33. If a single level of hazard protection is required, make sure to use containers that are tested to
Risk Group II or above and that more than one level of protection is provided in the case of
materials of a higher Risk Group.
WARNING
If the unit is not used as recommended by the manufacturer, the overall safety will be
impared.
WARNING
Check MSDS, wear required Personal Protective Equipment, and observe all applica-
ble local and national regulations before dispensing and disposing of hazardous sam-
ple.
WARNING
The laboratory procedure should ensure that no person or hazardous substances are
present within a 12" (30 cm) zone around the centrifuge when operating.
WARNING
This product does not contain bio-seals as per IEC/EN/CSA 61010-2-20 and cannot
provide any level of containment in case of a spill or release of toxic, radioactive, or
pathogenic micro-organisms thus these materials are not recommended to be used
in this product.
3