Thermo Scientific mySPIN 12 Manuel D'instructions page 10

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Trouble Shooting
Trouble
No power
present
Unit or display is not
operating normally
Excessive vibration
Excessive noise
Lid will not close
Lid will not open
WARNING
Spin balanced loads only!
Do not place the tubes asymmetrically nor load unequal volume tubes. Improper place-
ment of tubes will lead to insufficient centrifugation and may cause serious injury or an
accident.
Balancing the Rotor
Tubes of equal weight and size should be placed opposite each other. Use additional sample
or water in other tubes to provide a balanced rotor. Examples of proper sample balancing are
illustrated next.
Resolution
Verify that the AC adapter is fully plugged into the wall and
rear of unit.
Verify that the power switch is turned on.
Turn off the unit, wait 2 minutes, and turn power on.
Inspect the tubes for equal fill or improper placement.
Inspect the rotor for improper installation.
Remove the tubes and determine if the noise persists with an
empty rotor.
Verify that nothing is blocking the lid from fully closing.
Verify that nothing has fallen into the lock mechanism open-
ing.
If for any reason you need to manually open the lid to access tubes
due to an error or power loss, please perform the below:
Turn off the unit, remove power cord.
Make sure the rotor has stopped completely.
Use a thin rod and insert it into the opening on the bottom.
Press gently but firmly. You will feel a mechanical movement
within the unit and the lid will release.
Remove the rod, set the unit on the feet.
Remove your tubes and reclose the lid.
9

Publicité

Table des Matières
loading

Table des Matières