Xylem Ebro EBI 12 EX Serie Mode D'emploi page 19

Table des Matières

Publicité

English
data loggers with 3.0 mm flexible
probes
Do not bend the probe in the first 50 mm at
the housing nor in the 50 mm at the probe
tip.
This can damage the probe. If bent close to
the housing, the mechanical load is too great
due
to
the
edge
(figure 1).
The sensor is located in the 50 mm at the tip
of the probe.
Do not use pliers to bend it. When bending,
hold the probe tight in the lower area with two
fingers and bend it into the required position
with the other hand.
The bend radius must be at least 80 mm.
Smaller bend radii lead to material fatigue
and hence to breakage of the probe.
2/2021 BA80122def05_EBI 12 Serie-EX
of
the
housing
Français
enregistreurs de données avec
sondes souples de 3 mm
Ne pas plier la sonde, ni dans les 50
premiers millimètres au niveau du boîtier ni
dans les 50 mm de la pointe de la sonde.
Cela pourrait endommager la sonde. Pour le
pliage près du boîtier, la sollicitation
mécanique par l'arête du boîtier est trop
élevée (fig. 1).
C'est dans les 50 mm de la pointe de la
sonde que se trouve le capteur.
Ne pas utiliser de pince pour plier la sonde.
Pour le pliage, maintenir fermement la sonde
avec deux doigts dans la zone inférieure et
la plier avec l'autre main pour l'amener dans
la position désirée.
Le rayon de courbure doit être d'au moins 80
mm. Des rayons de courbure inférieurs
entraînent une fatigue de la matière et, de ce
fait, la rupture de la sonde.
19

Publicité

Table des Matières
loading

Table des Matières