Heathrow Scientific sprout Manuel D'instructions page 7

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Operation:
1.
Remove the contents from the package, examining them carefully for
breakage, defects, or missing parts. Save the packaging.
2.
Place Sprout
counter and other moving equipment. Make sure the power cord is
clear of hot surfaces and other hazards.
Warning!
Spin balanced loads only!
Do not place the tubes asymmetrically nor load unequal
volume tubes. Improper placement of tubes will lead to
insufficient centrifugation and may cause injury or accident.
3.
Verify that the power switch is in the OFF position. Plug the power
adapter cord into the rear of the unit and the power cord into an
approved outlet.
Turn power switch ON. If there is a smooth whirring sound
and the unit accelerates with little or no vibration, Sprout
use. If there are loud or unusual sounds, or excessive vibration, DO
NOT OPERATE. Contact Heathrow Scientific
4.
On benchtop prepare samples tubes in rack so that tubes are filled to
equal levels. Close tube lids.
5.
To open lid, firmly tilt lid toward rear of unit.
Warning!
The laboratory procedure should ensure that no person or hazardous
substances are present within a 12" (30 cm) zone around the
centrifuge when operating.
6.
Tubes of equal weight and size should be placed opposite each
other. Use sample or water in tube to provide balance. Proper sample
balancing illustrated below.
2 Tubes
®
on a sturdy, level surface away from the edge of the
SnapSpin
3 Tubes
Circular Rotor
4 Tubes
®
is ready to
®
LLC Technical Support.
6 Tubes
7

Publicité

Table des Matières
loading

Table des Matières