Lowenstein Medical Allegra M30 Mode D'emploi page 91

Masquer les pouces Voir aussi pour Allegra M30:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Emission limits
Conducted and radiated
CISPR 11
RF emissions
Harmonic distortion
IEC 61000-3-2
Voltage fluctu-
IEC 61000-3-3
ations/flicker
Housing
Phenomenon
Test method
Electrostatic discharge
IEC 61000-4-2
Radiated RF
IEC 61000-4-3
disturbances
Magnetic field at
IEC 61000-4-8
rated power
frequency (50 Hz)
Immunity test levels
±8 kV contact
±15 kV air
28 V/m
80 MHz to 6 GHz
80 % AM at 1 kHz
30 A/m
Power supply AC
Phenomenon
Test method
Fast transient electrical
IEC 61000-4-4
disturbances/bursts
Impulse voltages/surges
IEC 61000-4-5
Conducted RF distur-
IEC 61000-4-6
bances
Voltage dips
IEC 61000-4-11
Voltage
IEC 61000-4-11
interruptions
Note: U
is the AC mains voltage prior to the application of the test levels.
T
EMC information
Immunity test levels
±2 kV
100 kHz repetition
frequency
±1 kV line-to-line
±2 kV line-to-ground
3 V
0.15 MHz to 80 MHz
6 V in ISM bands and amateur
radio bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
0 % U
for ½ period
T
At 0°, 45°, 90°, 135°, 180°,
225°, 270°, 315°
0 % U
for 1 period
T
70 % U
for 25 periods
T
one-phase at 0°
0 % U
for 250 cycles
T
EN
42