Troubleshooting; General Information; Visual Inspection; Checks With Multimeter - Sartorius PR 6201/15 Manuel D'installation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

8 Troubleshooting

8 Troubleshooting
8.1

General information

8.2

Visual inspection

Component
Weighing object
Junction box
Connecting cables
Mounting kit
Load cell
8.3

Checks with multimeter

8.3.1

Checking the zero output signal of the load cell

EN-41
The following hints will enable a technician to do an initial diagnostic or help in case
of incorrect or non-reproducible weighing results after commissioning and calibration.
Possible fault
Are all hoses, pipes and cables free of force shunts?
Are the connections flexible and coupled horizontally?
Are devices which are rigidly connected to the weighing system in direct contact
with the surrounding installation?
Is there friction between the weighing object and its surrounding (e.g. dirty
apertures, …)?
Has moisture penetrated into the interior?
Do all solder or screw connections provide safe contact?
Is the sheath damaged?
Has moisture penetrated into the interior?
Is there a contact between the lift-off protection and the scale itself?
Are the constrainers stuck?
Does the load cell stand vertically?
Is the measuring chamber cover damaged?
Is the sheath of the load cell cable damaged?
Has moisture penetrated the load cell cable?
-
Unload load cell.
-
Disconnect the load cell measuring outputs.
-
Check whether the output voltage without load is within the limits.
Type
L
D1/N/C3
for PR 6201/54..
LA
Output voltage
0 mV ±0.02 mV/V
0 mV ±0.01 mV/V
0 mV ±0.02 mV/V
3.2… 4 mA
GAIN connected to +supply voltage 24 ±4 V, see Chap-
ter 6.2.1.
Compression Load Cell PR 6201
Sartorius Intec

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Pr 6201/25Pr 6201/35

Table des Matières