4. Technical data
Model
Classification acc. to MPG
Operating voltage
Ambient operating
conditions
Sound pressure level
Power input
Coarse dust filter
Bacteria filter
Fuses
Interface
Weight
Dimensions (H x W x L)
Warranty
O2-concentration
Status display O2
(at operating temp.)
Max. recommended vol flow 6 l/min.
Volume flow (at 0 or 7kPa) 1 - 6 l/min. acc. to pre-settings
Max. output pressure
Oxyderme THBO (Kröber O2)
IIa
230 V, 50 Hz
Operation temperature: +5°C to +40°C
Storage temperature: -25°C to +70°C
Humidity: 15% - 93% r.H., non-condensing
Atmospheric pressure: 700 to 1,060 mbar
35 dB(A)
360 W
in back of housing
behind service flap
main side: 2 x TT2.5 AH 250 V
internal: 1 x T1.0 AL 250 V
USB
WARNING! Unsafe connections!
!
Connect only manufacturer-authorized devices to this USB
port.
19.8 kg
53.5 x 20.3 x 52 cm, without castors
30,000 hours, maximum 5 years
1 to 4 l/min. 95% - 3%
4 to 5 l/min. 85% +/- 3%
5 to 6 l/min. 75% +/- 3%
82% statuts display
60% deficiency display
70 kPa
11
Oxyderme THBO
(see chapter 1.4)