7. STERILISATION TABLE

CYCLES
Not wrapped
121° C
Not wrapped
134° C
The data of this table are indicative; the choice of the sterilisation cycle must be based on the data
supplied by the producer of the object you want to sterilise.
8. FILLING THE WATER TANK
Remove the water tank cap, fill with distilled or demineralised water up to reaching the lower edge of the
sheet separation wall (see dwrg. D).
NOTE: The use of feedwater with contaminants at levels exceeding those given in the table above
can greatly shorten the working life of the device, with damage at the components of the autoclave
and then invalidate the manufacturer's warranty.
TABLE OF QUALITY LEVEL IN ACCORDING TO DIRECTIVE DIN EN 285
CEN STANDARD DIN EN 285
Evaporation residue
Silicium oxide, SiO2
Iron
Cadmium
Lead
Rest of heavy metals except iron, cadmium, lead
Chloride (Cl')
Phosphate (P20s)
Conductivity (at 20° C)
PH value (degree of acidity)
Appearance
Hardness (E Ions of alkalin earth)
STERILISATION
DRYING TIME
TIME
20 min.
5 min.
10 min.
5 min.
MAX. LOAD for
WORKING PRESSURE
TRAY (Kg)
(Bar Relative)
1
1
dwrg D
10
1
0.2
0.005
0.05
0.1
2
0.5
15
5 to 7
Colourless clean without sediment
0.02
1,2÷1,3
2,1÷2,3
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
mg/l
µs/cm
mmol/l
16