  • FR

Les langues disponibles

-Sterilisation
Sterilisation in an autoclave conforming to: EN 13060,
EN ISO 17665-1
CAUTION!
- Follow the instructions of the autoclave manufacturer.
- Clean, disinfect and lubricate the handpiece before
sterilisation. The presence of blood residues on the
external or internal surfaces can lead to the formation
of corrosion.
- Make sure that the handpiece is properly dried after
sterilisation.
Authorised sterilisation procedures
1) Seal the handpiece and its accessories in sterilisation
bags conforming to EN 868-5.
2) Sterilise the bagged handpiece in an autoclave
according to EN ISO 17665-1 standard and respect the
following conditions:
- Class B steam sterilisation in autoclaves conforming to
54
Eng
EN 13060 standard. Sterilisation time at least 3.5 minutes
at 135 °C.
3) Remove the bag from the autoclave.
-Storage
- Keep the handpiece in the sterilisation bag until it is
needed.
- Once sterilised, store the product in a dry and dust-free
place, if possible with a low microbial charge.
CAUTION!
Respect the validity date for the sterilisation!

Disposal

Sterilise the product and its accessories before
disposing of them. Refer the disposal of the product
to authorised companies qualified to dispose of
special industrial waste.