Calibration - OHAUS EB Serie Manuel D'instruction

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

6. CALIBRATION

For best results, calibrate the scale at regular intervals. Temperature changes,
gravity variations, altitude changes and abuse are few reasons why a scale may
need recalibration.
When the scale is ideally positioned for operation, enter calibration and proceed as
follows:
Long-press ZERO and COUNT at the same time to initiate
calibration. The required calibration weight flashes on the
display (ex. 3000g).
At this time, a different calibration weight value can be selected by pressing
Available selections are 1/3, 2/3 and 100% of full capacity.
Once the desired value is displayed, place the corresponding weight on the pan.
The display flashes until the actual weight is registered and the calibration process
ends (the scale re-starts). At this time, remove the weight on the pan before the
display count-down sequence ends.
NOTES:
Calibration weight is displayed in grams only.
To abort the calibration process, turn the scale off.
If needed, the factory default calibration value can be recovered. As the
scale is turned on, press TARE and COUNT at the same time during the power-on
sequence. The power-on sequence will re-cycle and the scale will reinitiate with the
default calibration.
EB SERIES EN-11
.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Eb3Eb6Eb15Eb30

Table des Matières