❏ When the test cycle is completed, the SDmatic makes three
"beeps".
The result screen is displayed.
The starch damage rate and two other calibration units are shown.
Result A
Result in UCD
❏ A
%: iodine absorption rate (%)
i
A
% = (1 - Ir/ Im) x 100
i
❏ UCD: the standard value depends on the standard conditions:
14% moisture content, 12% proteins.
❏ UCDc: value adjusted using the values of H2O% and prot%.
❏ AACC 76-31 / Farrand: customized units (refer to
The results can be printed by pressing on the icon "Print".
To leave this result screen, press on the icon "Exit".
SDmatic
%
I
A % : 95.59%
I
UCD : 25.1
AACC76-31
Farrand
Customized units
Result in UCDc
(adjusted)
UCD : 25.1
: 5.96
: 37.79
Print
Exit