YOKOGAWA Vigilantplant EXAxt SC450 Manuel D'instructions page 28

Transmetteur de conductivité ou de résistivité
Table des Matières

Publicité

Measurement setup
 Measure Conductivity only
Configure sensor
 Temperature settings
 Temp. Compensation
 Calibration settings
 Concentration
Enter
Note!
"Main parameter" et/ou "Measure" déterminent le reste de la structure des menus
Menu
Paramètre
Configure Sensor
Cell constant
Temp. Comp.
Reference Temp.
Manual Comp.
Manual Temp.
Temp. Coef
T.C.methods 1
Temp. Coef
T.C.methods 2
Conductivity only
Resistivity only
Conduct.+Concentr
Concentration only
Valeur
par défaut
min.
0.1 cm -1
0.005 cm -1
25ºC, 77ºF
0ºC, 32.0ºF
25ºC, 77ºF
-20ºC, -3.0ºF
2.10%/ºC, 1.17%/ºF
-10%/ºC, -0%/ºF
2.10%/ºC, 1.17%/ºF
0%/ºC, 0%/ºF
Etendue
max.
50 cm -1
100ºC, 211ºF
250ºC, 482ºF
3 . 5 % / º C ,
3.5%/ºC, 2%/ºF
IM 12D08N05-01F-E
21
2 % / º F