Maintenance
9.3

Errors/complications during the calibration

Action
Status display
Yellow/red flashes
alternately
Stop the calibration gas. Remove the calibration adaptor VarioGard 3x00.
Place the magnetic
wand on the
marking »
«
for the zero-point
Green flashes
calibration or on
alternately
the marking »
«
for the span
calibration for
2 seconds.
Remove the
Green flashes
magnetic wand.
alternately
Possible corrective measures:
Repeat the calibration.
Check the calibration gas concentration.
The flushing time of the calibration gas is not sufficient.
If there has been several unsuccessful attempts, replace the sensor.
48
Output signal Meaning
Error in
The device has
calibration
detected an error or
complications.
Maintenance
Error is
mode
acknowledged.
Maintenance
Measurement mode is
mode
activated with previous
setting for zero-point
or span.
9.4

Abort the calibration

Action
Status display
Stop the calibration gas. Remove the calibration adaptor VarioGard 3x00.
Place the magnetic
wand on the
marking »
«
for the zero-point
calibration or on
Yellow flashes
the marking »
«
for the span
calibration for
2 seconds.
Green/yellow
flashes alternately
Remove the
Green/yellow
magnetic wand.
flashes alternately
Output signal Meaning
The instrument has
Maintenance
detected an abort
mode
command from the
user.
Maintenance
Instrument
mode
acknowledges
the abort.
Maintenance
Measurement mode
mode
is activated with
previous setting for
zero-point or span.