Télécharger Imprimer la page

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 37

9. TROUBLESHOOTING

Warning
The table below lists some of the problems you may experience with your
humidifier and possible solutions.
Problem
The device reports a
mask leak but my
mask seems to be
fitting well.
If you notice any unexplained changes in the performance
of this device, if it is making unusual or harsh sounds, if it
has been dropped or mishandled, or if the enclosure is
broken, disconnect the mains cable from the therapy
device and discontinue use. Contact your healthcare
provider.
Possible Cause
The humidifier chamber is
not fully engaged or is
missing.
The silicone cover for air
outlet is missing from the
back of the device or not
fully inserted.
The gasket seal between
the two bowls of the water
chamber is missing.
TROUBLESHOOTING
Measure
Slide the humidifier chamber
fully into the cradle. Verify the
unlocking handle to the water
chamber clicks into place.
Make sure the silicone cover
for air outlet is fully inserted
into the back of the device.
Refer to Assembly in the
manual of therapy device for
instructions on system setup.
Insert the gasket seal into the
water chamber. Follow the
instructions for cleaning the
water chamber.
11

Publicité

loading