Dc Voltage Measurement / V - Testboy TB-313 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 43

DC voltage measurement / V=

Use the selector switch to set the appropriate range. Insert the black test
lead into the 'COM' socket and the red test lead into the V/Ω/TEMP/CAP
socket. Using the test probes, touch the test points of the test object. Read
measurement value on the display. The polarity of the voltage will also be
displayed.
DC voltage
Measurement range
200 mV
2 V
20 V
200 V
600 V
-Input resistance: 10 MΩ
Max. input voltage: 600 V DC
®
Testboy
TB-313
Resolution
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
Operation
Accuracy
± 0.5 % of reading + 1
digits
± 0.5 % of reading + 3
digits
± 0.8 % of reading + 3
digits
33