Fingernails and the luminescent tube should be on the same
side. Measured result. The values measured with this device
are not suitable for making diagnosis. The oxygen saturation
(SpO2) is a term referring to the concentration of oxygen in blood. The
value normally lies between 90% and 96%.
7. REPAIRING AND MAINTENANCE
• Change the batteries when the low voltage is displayed on the screen.
• Clean the surface of the device before using. With a soft cloth, moistened
with isopropyl alcohol, let it dry in air or clean it by dry clean fabric. To
prevent a cross infection from the previous use.
• Please take out the batteries if the device is not in use for a long time.
• The best storage environment of the device is -20° C to +55° C ambient
temperature and not higher than 93% relative humidity.
High-pressure sterilization cannot be used on the device.
Do not immerse the device in liquid. It is recommended that
the device should be kept in a dry environment. Humidity may
reduce the useful life of the device, or even damage it.
8. TROUBLESHOOTING
Trouble
Possible Reason
1. The finger is not
The SpO2 and pulse
positioned properly.
rate cannot be
2. The users SpO2 is
displayed normally
too low to be detected.
1. The finger is not
The SpO2 and
placed inside deep
pulse rate are not
enough.
displayed stably
2. The finger is shaking
or the user is moving.
1. The batteries are
drained or almost
drained.
The device cannot be
2. The batteries are not
turned on
inserted properly.
3. The malfunction of
the device.
1. The device will
enter standby mode
when no signal is in
The display is
the product within
off suddenly
5 seconds.
2. The batteries are
almost drained.

9. KEY OF SYMBOLS

Solution
1. Place the finger
properly and try again.
2. Try again.
1. Place the finger
properly and try again.
2. Let the patient keep
calm.
1. Change batteries.
2. Reinstall batteries.
3. Please contact the
local service center.
1. Normal.
2. Change batteries.
Symbol
Description
Type BF
Refer to instruction manual/booklet
The pulse oxygen saturation(%)
PRbpm
Pulse rate (bpm)
PI
Perfusion Index (%)
The battery voltage indication is deficient
(change the battery in time avoiding the
inexact measure)
1. No finger inserted
2. An indicator of signal inadequacy
Battery positive electrode
Battery cathode
1. Change brightness of the screen.
2. Exit standby mode.
SN
Serial number
WEEE (2002/96/EC)
IP22
International protection
This item is in according to the Medical Device
Directive 93/42/EEC of June 14, 1993, a directive of
the European Economic Community.
Manufacturer
Date of manufacture
Storage and Transport Temperature limitation
Storage and Transport Humidity limitation