Below are additional safety precautions that apply to
chromatography refrigerator models:
CAUTION: Equipment should only be powered
using the internal outlet. In case of a leak the
safety circuit will remove power to the outlet in
the unit, but shall have no control over equipment
power via battery or externally. Do not use
instrumentation or equipment that incorporates
potential ignition sources, e.g. open contact
switching, brushed DC and AC motors, etc.
CAUTION: The chromatography safety circuit
has slots / holes at the bottom of the plug panel
for enabling air to the safety sensor. Do not block
the Safety Circuit ventilation holes as this will
diminish and defeat the Safety Circuit.
WARNING: Do not store or use uncapped
reagents, vessels and bottles inside the
chromatography equipment. This may diminish
functionality of the safety circuit.
General Purpose and Chromatography Refrigerators
Safety Precautions |
3