Temperature Set point

The factory default temperature setpoint is 5°C for all
laboratory refrigerators. Adjusting the set point is not
recommended. In case the set point has to be changed,
follow the instructions given below.Set point can be adjusted
between 4 to 5.5°C only.
Press and release SET button and the controller will display
"SET". Press and release SET button again and the controller
will display setpoint.
Press Up / Down button to increase / decrease set point
temperature. Press and release the SET button to save the
new set point.
Table 5. Alarm Codes
Alarm
Trigger
code
"AH1"
Value read by Pb1
> HAL after time
of tAO.
(see 'MAX/MIN
TEMP. ALARMS ')
"AL1"
Value read by Pb1
< LAL after time of
tAO. (see 'MAX/
MIN TEMP.

ALARMS')

General Purpose and Chromatography Refrigerators
Automatic
Outputs
clearance
User
Registration AH1 label in the AL
Configured
folder
No effect on regulation
Label AH1 displayed
alternately with the actual
value read by the probe
Pb1
User
Registration AL1 label in the AL
Configured
folder
• No effect on regulation
• Label AL1 displayed
alternately with the actual value
read by the probe Pb1
The display will automatically reverts to current temperature
after 30seconds, if SET button is not pressed.
Alarms
The alarm system is designed to provide visual and audible
warning signals for both power failure and rise in temperature.
The alarm is equipped with a battery backup.
The power failure alarm system is activated only when
the key switch is turned to the Power Failure Alarm ON
position.
Low temperature, high temperature and door ajar alarms are
set to ON position by default as soon as the controller is
turned on. These cannot be turned off.
Default low and high alarm values are 2°C and 10°C. These
values may be adjusted, following instructions in section
Temperature Set point.
Note: There is no door open alarm on sliding door model.
The audible warning signal sounds when there is a power
failure, temperature alarm condition, or when the door is ajar
for more than 1 minute.
Alarm display codes for high temperature, low temperature,
sensor failure, door ajar are mentioned in Table 5.