Enter the value by pressing the PUMP CURRENT
pushbutton.
Press the CYCLE NUMBER pushbutton to select the
second number.
S P E E D
T H R E S H O L D
S E L E C T
I O N : X X K R P M
Enter the value by pressing the PUMP CURRENT
push-button, and the display shows:
R U N - U P
S E L
:
X X h
where: RUN-UP TIME = is the interval time from
start to speed threshold value in hours, minutes,
seconds. Select from 00 to 99 hours, and from 00 to
59 minutes or seconds.
Select the run-up time according to the chamber
volume and/or operating cycle feature (see the
preceeding cycle diagram) by pressing the CYCLE
NUMBER pushbutton to select the desired number.
64
TECHNICAL INFORMATION
T
I M E
X X m
X X s
87-900-854-01 (C)