The PosiTector SST can be adjusted to a known conductivity
standard. Use the (-)(+) buttons to decrease/increase the
displayed value to match the known conductivity of the standard.
NOTE:
In typical use, a calibration adjustment should not be
required. Before performing the adjustment, ensure that the
PosiTector SST test cell has been thoroughly cleaned with
DI water. Ensure that the conductivity standard and
PosiTector SST test cell are at the same temperature.
When checked (default), the Gage guides the user through a 3-step
test process in accordance with ISO 8502-6. Uncheck Bresle Mode
to use the PosiTector SST as a regular conductivity meter such as
when testing abrasive materials in accordance with ISO 11127-6
and ASTM D4940-15.
Hi Res
Increases the displayed Gage resolution as follows:
Gage accuracy is not affected.
NOTE:
Battery Type
Selects the type of batteries used in the Gage from a choice of
"Alkaline", "Lithium" or "NiMH" (nickel-metal hydride rechargeable).
If "NiMH" is selected, the Gage will trickle charge the batteries while
connected via USB to a PC or optional AC power supply (USBAC).
The battery state indicator symbol is calibrated for the selected
battery type. No damage will occur if the battery type used in the
Gage does not match the selected battery type.
The PosiTector SST has internal memory storage for recording
measurement data. Stored measurements can be reviewed on-
screen or accessed via computers, tablets and smart phones.
All stored measurements are date and time-stamped.
The
symbol appears when the Gage is set to store
measurement data.
(Calibration Adjustment)
0.1 µS, 0 - 200 µS
1.0 µS, over 200 µS