Digisystem, Functional Checks
Checking the
performance
The purpose of the following procedure is to verify the performance of the Transmitter. It is
important that the test is conducted away from areas of electromagnetic interference.
1. Plug the Transmitter's cable set into the connection socket.
2. Connect the black and red cable clips together, ensuring good metal to metal contact.
3. Hold down Frequency Control Button and switch the Transmitter on. Note, Hold the
frequency control button down until the Extended Self Test has commenced.
4. Observe the displayed output:
Induction Mode Test
Connection Mode Test
5. The transmitter will display the outcome.
Pass
Battery indicator:
Audible Output:
Fail
Battery indicator:
Audible Output:
Mode Display:
Frequency Display:
Mode Display:
Induction LED illuminated.
Frequency Display: Frequency LED illuminates showing the
frequency on test.
Mode Display:
Induction LED off.
Mode Display:
Connection LED illuminated.
Frequency Display: Frequency LED's illuminates showing the
frequency on test.
Mode Display:
Connection LED off.
LED Flash.
High - Low pulsed tone emitted three times.
LED Flash.
Low pitch tone emitted.
Induction or Connection LED illuminated to show failure
mode.
Frequency LED illuminated to show frequency failure.