OPERATING TEST
To check that the device works, after finishing the installation operations, switch on at the autoclave master
switch and activate the program for water loading from the PURA™ (see instructions given in the autoclave
manual).
If the led that indicates the max level of water in the autoclave tank is off, the PURA™ will call for water.
When the max level is reached, the related led of the autoclave comes on, delivery of water stops and the
green led of the PURA™ goes out.
If the autoclave tank is at the max level, the water must be drained out by opening the tap at the back of the
machine in order to check the operation of the PURA™.
When the max level led of the autoclave goes out, delivery of water from the PURA™ starts and the green
led comes on. Delivery of water from the PURA™ continues until the max level is reached.
9

OPERATORS ISTRUCTIONS

CONNECT PURA™ ONLY TO AUTOCLAVES COMPATIBLE WITH THAT UNIT.
ALWAYS INSTALL A TAP UPSTREAM OF THE DEMINERALIZATION CIRCUIT IN ORDER TO AVOID
FLOODS;
AT THE END OF THE DAY OR WHEN THE AUTOCLAVE IS NOT GOING TO BE USED FOR SOME
DAYS, CLOSE THE TAP INSTALLED UPSTREAM OF PURA™
CONNECTING PURA™ TO THE HOT WATER LINE IS STRICTLY FORBIDDEN!
PURA™ CAN BE MOUNTED TO THE WALL OR CAN BE SUPPORTED ON A SURFACE.
THERE IS NO MINIMUM CAPACITY. THE MAXIMUM CAPACITY IS 12 l/h
MAXIMUM AND MINIMUM PRESSURE FOR WATER CLEAN OPERATION: from 2 to 7 bar
USE OF PURA™ IS GRENTED WITH MAX HARDNESS OF THE INCOMING WATER OF 200 °F.
OUTPUT CONDUCTIVITY VALUE: < 15 µS (Micro Siemens); THE INPUT HARDNESS IS IRRELEVANT.
THE MEMBRANE AND RESINS NEUTRALIZE ALL KINDS OF ORGANIC AND INORGANIC SALT.
THE CARTRIDGES MUST BE HANDLED WITH NORMAL PRECAUTIONS. THE USE OF PROTECTIVE
GLOVES IS MANDATORY
THE FILTERING CARTRIDGES MUST BE DISPOSED OF AS SPECIAL WASTE (SEE SEC.7)
INSTALL PURA™ WITHIN 2 METRE OF THE AUTOCLAVE
DO NOT EXPOSE PURA™ THE WATER CLEAN TO FROST
DO NOT INSTALL NEAR SOURCES OF HEAT
DO NOT INSTALL UNDER OTHER TAPS