respecting the manufacturer's Warnings
and precisely following its Instruction for
Use.
Disinfectant shall be poured into a
proper tray having secure lid. Tray size
shall
have
to
submersion of the device and to facilitate
the evacuation of air from the inside.
Warm up the disinfectant to at least
20°C.
Immerse the product completely, and if
applicable, fill all lumens and eliminate air
pockets by means of a syringe fulfilled of
disinfectant (to guaranty full internal
lumen contact with the disinfectant).
Close the tray with the secure lid.
Wait for at least 5 minutes and be sure
that the temperature of the disinfectant
solution doesn't drop below 20°C.
The
ASP
CIDEX
requires a total of three (3) rinses, with
large volumes of fresh water to properly
remove its residues. Sterile water shall
be poured into proper rinsing trays. Tray
size shall have to allow the complete
submersion of the device and to facilitate
the evacuation of air from the inside.
For each of the 3 rinses do:
- Flush all lumens with large volumes
(not less than 100 ml) of rinse water.
Use a syringe or a water jet pistol to
correctly accomplish this task.
- Keep the device totally immersed for
a minimum of 1 minute in duration.
- Remove the device from the rinsing
tray and discard the rinse water.
Finally, fully dry it internally and externally.
Always use fresh volumes of water for
each rinse. Do not reuse the water for rinsing
or
any
other
purpose.
disinfectant may cause serious side effects.
ENGLISH
allow
the
complete
OPA
disinfectant
Residues
FB-618/EN – Rev. A-01 - 2017.03.30
The following validated (on Miele G 7735 CD
using program "DESIN 93° C-3'") process can
be used to achieve an A
Holding time: 3 minutes
Min temperature: 93°C
Final automated drying cycle executed in full
Special instructions and warnings of the
manufacturer of the washer disinfector have to
be followed.
Follow carefully the instructions provided
by the disinfection solution manufacturer.
of
Level of 3000:
0
29/46