Failure
Transmitter does not go
into "HOLD" mode
Inductive check-back position indica-
tor not recognized by EasyClean
Screen is blank
Alarm message "Check Levels"
appears on the display
Error message
TEST EEPROM
TEST DISPLAY
TEST CLOCK
ERROR 12C
ERROR: EEPROM
ERROR : CLOCK
ERROR: LCD
ERROR: ACLOW
02/2021 © METTLER TOLEDO
Printed in Switzerland
Cause
Terminal contact dry.
"HOLD" or "CONTROL" signal
does not trigger.
Inductive check-back position
indicator defective.
Terminal contact dry.
LCD screen defective.
Power supply interrupted.
The flow sensor cannot identify any
media flow during a rinsing and/or
cleaning cycle.
Terminal contact dry.
Error messages (ERROR) on the programmable control
Cause
Self-test interrupted
Control defective
Control circuit diagram storage is
defective
Clock error
LCD defective
Control defective
EasyClean 200e
Remedy
Check the wiring as described in
section 4.2 "Connections".
Check the wiring on the transmitter.
Check whether green LED on auxiliary
print in control unit lights up for 5 sec –
Pause – 2 sec on changeover from "RUN"
to "SERVICE". Possibly replace the print.
Replace inductive check-back
position indicator.
Check the wiring as described in
section 4.2. Check cable.
Replace the print unit with the PLC.
Check the power supply.
Check flow sensor.
Check water or detergent valve.
Check pipe.
Check the wiring as described in
section 4.2. Check cable.
Remedy
Replace the print unit with the PLC.
Replace the print unit with the PLC.
Replace the print unit with the PLC.
Replace the print unit with the PLC.
Replace the print unit with the PLC.
Replace the print unit with the PLC.
EasyClean 200e
30 404 676 F