Program sequence
02/2021 © METTLER TOLEDO
Printed in Switzerland
1
Calculation:
T = (Tube length [m]
12.5*
Example:
T = (5
12.5
30)
23
s, C3 and C4 have to have at least this value.
4
Viscosity and tube length influence the delivery rate of the pump. The maximum delivery rate of the diaphragm pump
used is 5 ml/s. Viscosities above 25 Pa s and tube lengths greater than 10 m (32.8 ft) can reduce the rate to 3 ml/s.
We calculate with an average of 4 ml/s. To ensure complete filling of the flushing chamber, check the outflow of the
chamber (dismount drain water hose), and check the outflow of cleaning agent after delivery. If necessary readjust
parameter C3.
Flushing chamber capacity overview:
Type
Capacity
InTrac 7XX single chamber
20 ml
InTrac 7XX double chamber 50 ml
InTrac 798e
70 ml
2 Table for tube length and appropriate
transport time:
Tube length
C5
1...5 m (3...17 ft)
4 sec.
5...8 m (17...26 ft)
6 sec.
8...10 m (26...33 ft)
8 sec.
The EasyClean is supplied with standard program sequences. Between the individual functions
and always at the end of a process, the whole system is rinsed with water and air.
The standard settings result in the following program cycle times:
Cleaning: approx. 5 minutes
flushing chamber capacity [ml])
4 [ml/s]
EasyClean 200e
* fix, tube capacity per m
EasyClean 200e