Immunity test
IEC 60601 test level
Power frequency
(50/60 Hz)
magnetic field
3 A/m
IEC 61000-4-8
NOTE
U
is the a. c. mains voltage prior to application of the test level.
T
Conducted RF
3 V
IEC 61000-4-6
Radiated RF
3 V/m
IEC 61000-4-3
Compliance level
3 A/m
3 V
3 V/m
29
Electromagnetic environment - guidance
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
Portable and mobile RF communications equipment
should be used no closer to any part of the
EQUIPMENT medical bed Medley Ergo including cables,
than the recommended separation distance calculated
from the equation applicable to the frequency of the
transmitter.
Recommended separation distance
3
5 ,
=
d
[
]
P
V
1
3
5 ,
80 MHz to 800 MHz
=
d
[
]
P
E
1
7
800 MHz to 2,5 GHz
=
d
[
]
P
E
1