To Install The New Seals And O-Rings - Agilent Technologies TPS-compact Mode D'emploi

Masquer les pouces Voir aussi pour TPS-compact:
Table des Matières

Publicité

17
Technical Information
Service Options
5

To Install the new Seals and O-Rings

1
2
3
4
5
6
7
8
9
10 Disable the "TIP-SEAL SERV" warning message by serial
304/314
If compressed air is available, blow any remaining seal debris off
the scroll parts. If seal debris is attached to the sides, use a razor
or Exacto knife to scrape the debris off.
Unpack the Tip-Seals.
Two tip seals are provided in the kit; one fits the orbiting scroll
and the other fits the out-board housing scroll.
Install the correct tip seal into the groove of the orbiting scroll by:
a
Facing the white plastic surface upwards and inserting the
foam backing into the groove. Start from the center and work
outward.
b
Cutting the tip seal off about 1/8" (3 mm) from the outer end
of the groove.
Install the correct tip seal into the groove of the outboard housing
scroll as in step 3.
Place the new main O-ring into the groove in the frame.
Ensure that the groove is clean.
Ensure the sealing face of the outboard hous-ing is clean.
Carefully replace the outboard housing by lining up the locating
pins. Ensure the tip seal has not fallen out of its groove.
Reinstall (4) M5 bolts (B) and torque the (4) M5 bolts to 5.6 Nm
(50 in-lb).
Unlock the upper part of TPS-compact by sliding the small lever
you can find on the bottom side of TPS-compact.
Reconnect the TPS-compact to the electrical power mains.
interface. To disable the Tip-seal Service message you have to
reset the Tip-seal hour-meter (Win 358). Write a "0" on the
window 358. The message will be showed again after 5000 h of
operation. To check the Tip-seal life you have just to read the
Win 358.