Getting Started; Features And Accessories - Tektronix P6246 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Getting Started

The P6246 and P6247 are high-bandwidth differential probes with
the TekProbe II interface. The P6246 probe has a bandwidth of
400 MHz and the P6247 has a bandwidth of 1 GHz. Both probes
have low circuit loading, high common-mode rejection, and come
with accessories that accommodate a wide variety of applications.
The TekProbe II interface is standard on Tektronix TDS Series
oscilloscopes. The Tektronix 1103 TekProbe Power Supply can be
used for instruments without the TekProbe interface (refer to
page 11).

Features and Accessories

Table 1 shows the features and standard accessories of the P6246 and
P6247 differential probes.
Table 1: P6246 and P6247 features and standard accessories
Feature/Accessory
P6246 & P6247 Instructions
Description
TekProbe II interface. The TekProbe II interface supplies power
to the probe, selects the correct display scaling, and automatically
sets the 50 Ω termination on the oscilloscope input.
If your oscilloscope does not have the TekProbe II interface, you
can use the optional 1103 power supply (refer to page 11).
Bandwidth limit. The bandwidth limit switch selects either the
specified bandwidth of the probe or limits the bandwidth to
200 MHz.
Probe attenuation. The attenuation switch selects either divide
by 1 (÷ 1) or divide by 10 (÷ 10).
7

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

P6247

Table des Matières