Measured values vary all the time.
???
Mental and physical condition or measurement condition is influencing blood
pressure. Make measurement under the same conditions.
Measured value is different from that taken at a hospital.
???
Mental condition such as nervousness at a hospital or relaxed feeling at home is
influencing blood pressure. Take records of blood pressure taken at home
and consult with your doctor.
Blood pressure is changing 24 hours a day in relation with mental and physical
conditions. Eating, drinking, smoking, exercising or bathing will alter blood pressure.
Blood pressure will also be influenced by mental condition such as feeling tense or
relieved and health condition. It is important to know the trend of blood pressure by
making measurement regularly. Make your own rules including conditions and time to
make measurement and take blood pressure everyday.
Inflation is repeated.
???
The cuff is inflated again if sudden change of pressure, which could be due to
movement or straining of arm or wrist, was detected. Inflation is repeated until
blood pressure is taken. Repeated inflation is not a malfunction of the
monitor. Do not move or speak during measurement.
is displayed.
Over pressure: blood pressure could not be taken due to moving or speaking
although the cuff was inflated to the maximum pressure. Do not move or speak
during measurement.
or
is displayed.
Blood pressure could not be taken due to moving or speaking. Do not move or
speak during measurement.
is displayed.
The cuff is not securely connected to the monitor. Secure the air connector.
The cuff is not wrapped appropriately. Reconfirm how to put the cuff on your arm.
is displayed.
Batteries are weak. Replace all batteries with new ones.
Nothing is displayed.
???
Batteries are depleted. Replace all batteries with new ones.
Batteries are inserted incorrectly. Reinsert the batteries correctly.
Battery terminals are not clean. Clean the terminals with dry cloth.
AC adaptor is not connected. Connect AC adaptor.
Touching the keys with tip of finger or nail is touching the key. Touch the key
with the plane of your finger.
Clock is not displayed.
???
The clock is inactivated. NOTE: Clock is inactivated when batteries and/or AC
adaptor are/is removed. Adjust date and time and activate the clock.
Measurement date and time are displayed with "
???
The clock is not activated. Adjust date and time and activate the clock.
Measurement was taken before the clock was activated. Measurement date
and time cannot be saved without clock being activated.
is displayed.
"START/STOP" KEY was being touched when batteries are being installed or
when AC adaptor is being plugged in. Turn off the monitor once with
"START/STOP" KEY.
If you cannot get correct measurement with the methods above, contact your dealer. Do
not disassemble or tamper with the internal mechanism.