Guidance and manufacturer's declaration – electromagnetic immunity
The device is inteded for use in the electromagnetic environment specified below. The customer or the user of the
device should assure that is used in suche environment
Immunity test
Test level EN
60601-1-2
Conducted RF
3 Vrms 150kHz to
80MHz
EN 61000-4-6
RF Radiata
3 Vrms 80MHz to
2,5GHz
EN 61000-4-3
Field strangths from fixed RF transmitters, are determined by an electromagnetic site survey, should be less than the
complicance level in each frequency rage.
Interference may occur in the vicinity of equipment marked with the following symbol:
Recommended separation distances between portable and mobile communications equipment and the
device
The device is intended for the use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the device can help prevent electromagnetic interferences by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the device as
recommended below, according to the maximum output power of the communication equipment.
Rated maximum power
of the transmitter (W)
0,01
0,1
1
10
100
IACER S.r.l.
Compliance level
3 Vrms 150kHz to
80MHz
3 Vrms 80MHz to
2,5GHz
Separation distance according to the frequency of the transmitter (m)
150kHz to 80MHz
d = 1,2 P
0,12
0,38
1,2
3,8
12
33 di 36
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance:
d = 1,2 P 150kHz to 80MHz
d = 1,2 P 80 MHz to 800 MHz
d = 2,3 P 800 MHz to 2,5 GHz
where (P) is the maximum output power
rating of the transmitter in Watts (W)
according to the transmitter manufacturer and
(d) is the recommended separation distance in
metres (m).
80MHz to 800MHz
800MHz to 2GHz
d = 1,2 P
d = 2,3 P
0,12
0,23
0,38
0,73
1,2
2,3
3,8
7,3
12
23